1 / 5
next product »»
|
XR440 Data Logger
|
|
The XR440 Pocket Logger is a 4 channel pocket size data logger suitable for monitoring temperature, humidity, light, AC current and voltage, Dc voltages, 4 - 20mA signals. Low power design makes this ideal for remote installations. Logging rates are programmable from 200Hz to 12 hours. Memory up to 129,000 readings, non volatile. The resolution is selectable from 8 to 12 bit. Supplied with free Windows software with real time graphs, simple point and click programming, Excel export facility and modem support. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Features
- Four multi-purpose inputs; mix and match PACE sensors in any combination.
- Each input will also read resistance, 0-5 vdc or a contact closure.
- Data retention of over 200 years with no power.
- High accuracy, up to +/-0.15°C and +/-0.25% F.S. with no calibration.
- Adjustable resolution of 12, 10 or 8 bits (lowering resolution increases reading storage capacity, see specifications).
- Wide operating temperature range -40°C to 60°C (-40°F to 140°F).
- Certificate of Validation, traceable to NIST, included with each unit.
- Long battery life, runs unattended for over 2 years (including sensors) without external power.
- Optional Pulse Module for tipping bucket rain gage, gas and water meters, and kWH meters.
- Optional Scaling Module for +/-5, 0-10, ±17, 0-20, ±27, 0-30, ±60, 0-60 vdc and 4-20 ma signals.
- Optional Thermocouple / Millivolt Modules for thermocouple or millivolt signal.
- Windows based software (full version available for free download) for logger setup, data transfer, graphing and data export.
The XR440 Data Logger was designed for easy setup and trouble free operation. Direct-connect sensors include Temperature, Humidity, Pressure, AC current, AC Voltage, Light and Process Signals. The Data Logger's four input channels accept any combination of the above sensors, plus resistive, 0-5 vdc signals, and contact closures.
Stand alone: The XR440 Data Logger does not require a computer while it is logging. You connect it to a computer using a serial RS232 cable (Part No. IC209, listed below) for setup, data transfer or to display readings in real time. You can also communicate with the Data Logger using cellular or land-line modems, or a wireless link.
Long life: Battery life with any combination of sensors is over two years (continuous operation). All sensors are powered from the logger's battery. The battery is user-replaceable (standard 9 volt size). |